Product features
Product features
Compliance with communications, medical safety certification
2 MOPP isolation protection
Leakage current ≤ 100μA
Level 6 energy efficiency
Standby power consumption ≤0.075W
12.0 or 24.0V output, up to 36W
Working sea dial height 5000 meters
Input Characteristics
Input voltage 90-264VAC
Frequency range 47-63 Hz
Input current 0.5A, 90VAC input condition
Surge current 100A Max at 240VAC input and cold start conditions
Contact current ≤100 μa at 264VAC input
The environment In general
Working temperature of 0 ~ 45 ℃ overall dimensions (length) 84.95 x51 (wide) x27.7 (high) mm
Storage temperature -20 ~ 60℃ Weight 155g
Operating humidity 10-90% RH, non-condensing mean time between failures & GT; 100,000 hours MIL-HDBK-217 at 25 ° C ambient temperature.
Storage humidity 5-90% RH
The working altitude is 5000m
Safety Approval
Safety/organization Medica I.T.E
CCC - GB4943.1
CB IEC60601-1 -
TuV Rheinland/Mark EN60601-1 -
FCC - FCC PART 15
NRTL NRTL AAMIES60601 -
Models & Parameters
Model | volgage | electric current | Rated power | Ripple and noise | Voltage tolerance | Linearity and load | Average efficiency | Start delay time |
(V) | (A) | (max) | Adjustment rate | |||||
UES36WZ1-XXXYYYSPA | 12.0 | 0.01-3.00 | 36.0W | 120mVpk-pk | ±5% | Linear adjustment rate: ±1% Load adjustment rate: ±5% | 88.30% | ≤3s |
19.0 | 0.01-1.89 | 36.0W | 240mVpk-pk | ±5% | 88.30% | ≤3s | ||
24.0 | 0.01-1.50 | 36.0W | 240mVpk-pk | ±5% | 88.30% | ≤3s |
landline