Product features
Product features
Meets medical safety certification
2 MOPP isolation protection
Leakage current ≤ 100μA
Level 6 energy efficiency
Standby power consumption ≤0.5W
Peak load: 120% of rated load (1s)
Working sea dial height 5000 meters
Dip switches control the output
No wind design features natural air cooling
Input Characteristics
Input voltage 90-264VAC(90-115VAC reference derating curve)
The frequency range is 47-63Hz
Input current 4.0a at 100VAC input
The environment general
Working temperature - 30 ~ 70 ℃ (50-70 ℃ reference derating curve) dimension 127 long) x75.6 (wide) x34.0 (high) mm
Storage temperature - 40 ~ 85 ℃ 300 g in weight
Working humidity 5 ~ 90% RH, noncondensable MTBF & gt; 200,000 HRS MIL-HDBK-217 at 25 ° C ambient temperature.
Storage humidity 5-95% RH
The working altitude is 5000m
Safety Approval
Safety organization/sign medical (meet) communication (meet)
CB IEC60601-1 IEC60950-1 IEC62368-1
UL ANSI/AAMI ES60601-1 CAN/CSA C22.2 NO.60601-1
UL60950-1 UL62368-1
TUV-SUD Mark EN60601-1 -
CE EN62368-1
Models & Parameters
Model | voltage (V) | electric current(A) | power (W) | Ripple (maximum) mVpk-pk | Voltage tolerance | Linear and negative adjustment rates | Average efficiency | Start delay time |
UES300-XXXYYY-OP | 12.00 | 0.01-22.5 | 270 | 120 | ±5% | 92% | ≤3s | |
15.00 | 0.01-20.0 | 300 | 120 | Linear adjustment rate: ±1% | 92% | ≤3s | ||
18.00 | 0.01-16.66 | 300 | 150 | 93% | ≤3s | |||
24.00 | 0.01-12.5 | 300 | 150 | Load adjustment rate: ±5% | 93% | ≤3s | ||
36.00 | 0.01-8.33 | 300 | 200 | 94% | ≤3s | |||
48.00 | 0.01-6.25 | 300 | 250 | 94% | ≤3s |
landline