Product features
Product features
Meets medical safety certification
2 MOPP isolation protection
Leakage current ≤ 100μA
Level 6 energy efficiency
Standby power consumption ≤0.21W
Peak load: 200% rated load @230Vac
Working sea dial height 5000 meters
Dip switches control the output
No wind design features natural air cooling
Input Characteristics
Input voltage 90-264VAC(90-115VAC reference derating curve)
The frequency range is 47-63Hz
Input current 1.5A, 100VAC input condition
Environment Commonly
Working temperature - 30 ~ 70 ℃ (45-70 ℃ reference derating curve) dimension 101.6 long) x50.8 (wide) x29.0 (high) mm
Storage temperature - 40 ~ 80 ℃ 190 g in weight
Working humidity 5 ~ 90% RH, noncondensable MTBF & gt; 200,000 HRS MIL-HDBK-217 at 25 ° C ambient temperature.
Storage humidity 5-95% RH
The working altitude is 5000m
Safety Approval |
Safety organization/sign Medical I.T.E |
CB IEC60601-1 - |
TUV-SUD Mark EN60601-1 - |
NRTL ANSI/AAMI ES60601-1 - |
Models & Parameters
Model | Voltage(V) | electric current(A) | power(W) | Ripple (maximum)mVpk-pk | Voltage tolerance | Linear and negative adjustment rates | Average efficiency | Start delay time |
UES90-XXXYYYSPA-OP | 12.00 | 0.01-7.00 | 84.0 | 100 | ±5% | 89% | ≤3s | |
15.00 | 0.01-6.00 | 90.0 | 120 | Linear adjustment rate: ±1% | 89% | ≤3s | ||
18.00 | 0.01-5.00 | 90.0 | 150 | 89% | ≤3s | |||
24.00 | 0.01-3.75 | 90.0 | 150 | Load adjustment rate: ±5% | 90% | ≤3s | ||
36.00 | 0.01-2.50 | 90.0 | 150 | 90% | ≤3s | |||
48.00 | 0.01-1.87 | 90.0 | 150 | 90% | ≤3s |