Product features
Product features
Compliance with medical safety certification
2 MOPP isolation protection
Leakage current ≤ 100 μ A
Comply with class VI energy efficiency
Standby power consumption ≤ 0.5W
Peak load: 120% rated load (1s)
Working altitude 5000 meters
Dial switch control output
No fan cooling, natural cooling
Input Characteristics
Input voltage 90-264VAC (90-115vac reference derating curve)
Frequency range 47-63Hz
Input current 3.0A, under 100vac input condition
Environment commonly
Working temperature -30~70 ℃ (50- 70 ℃ refer to derating curve) overall dimension 127.0 (length) x75.6 (width) x34.0 (height) mm
Storage temperature -40~85 ℃, weight 300g
Working humidity 5~90% RH, non condensing mean time between failures >200000hrs mil-hdbk-217, at an ambient temperature of 25 ℃.
Storage humidity 5~95% RH
Working altitude 5000m
Safety regulation certification
Safety regulation organization/sign Medical (compliant) Communication (compliant)
CB IEC60601-1 IEC60950-1 IEC62368-1
UL ANSI/AAMI ES60601-1
CAN/CSA C22.2 NO.60601-1 UL60950-1 UL62368-1
TUV-SUD Mark EN60601-1 -
CE EN62368-1
Models & Parameters
Model | Model voltage (V) | electric current(A) | power(W) | Ripple (maximum) mVpk-pk | Voltage tolerance | Linear and negative adjustment rates | Average efficiency | Start delay time |
UES240-XXXYYYSPA-OP | 12.00 | 0.01-18.00 | 216 | 120 | ±5% | 92% | ≤3s | |
15.00 | 0.01-14.4 | 216 | 120 | Linear adjustment rate: ±1% | 92% | ≤3s | ||
18.00 | 0.01-13.3 | 240 | 150 | 93% | ≤3s | |||
24.00 | 0.01-10.0 | 240 | 150 | Load adjustment rate: ±5% | 93% | ≤3s | ||
36.00 | 0.01-6.6 | 240 | 200 | 94% | ≤3s | |||
48.00 | 0.01-5.0 | 240 | 250 | 94% | ≤3s |
landline